products
services
  • Products

  • Manufacturer

  • Ambulatory Infusion Pumps
  • Pole Mounted Infusion Pumps
  • Syringe Infusion Pumps
  • Enteral Pumps
  • Elastomeric Pumps
  • Disposables
  • CPM Machines

CADD Prizm 6100 and 6101 For continuous, clinician activated bolus, and patient activated bolus infusions.

Recommended uses: Intravenous, Intra-arterial, intraperitoneal, epidural, or sub-arachnoid infusion of PCA or Continuous
Rate parameters: 0 - 30 ml/hr (or mg or m g equivalent)
Reservoir volume: 1 - 9999 ml. programmable in 1 ml increments, displayed in 0.1 ml incr.
Demand dose (PCA): 0 - 9.9 ml. in 0.1 ml increments (or mg or m g equivalent)delivers at approx. 125 ml/hr.
Battery Life: 9-volt alkaline, 12 hours at 100 ml/hr approximate
Occlusion Pressure: 18 ± 9 PSI
Accuracy: ± 6 %
CADD Prizm VIP 6100/6101:
Recommended uses: Intravenous, Intra-arterial, intraperitoneal, epidural, or subarachnoid infusion of PCA, TPN, Continuous, or Intermittent medications.
Rate parameters: Continuous mode: 0.1 - 350 ml/hr. in 0.1 ml increments from 0.1 - 100 ml/hr in 1 ml increments from 100 - 350 ml/hr (or mg or m g equivalent) PCA mode: 0 - 30 ml/hr (or mg or m g equivalent), TPN mode: 10.0 - 350 ml/hr in 10 ml increments
Battery Life: 9-volt alkaline, 12 hours at 100 ml/hr approximate. For EPS specs, see PDCADDTPN
Occlusion Pressure: 18 ± 9 PSI
Accuracy: ± 6 %
Pump Resolution: Standard reservoir or set = 0.050 ml/pump stroke nominal High volume set = 0.100 ml/pump stroke nominal
Dimensions: 1.7" D x 4.1" W x 5.6" H excluding cassette
Weight: 20 oz including 9 volt battery and empty 100 ml reservoir
Power Source: 9 volt alkaline or lithium battery, External Power Source, or AC Adapter

Price: $0.00